Disposable electrodes for ECG, in liquid GEL with Ag/AgCl (Silver/Silver Chloride) sensor and stainless steel clip.

FIAB disposable ECG electrodes are suitable for recording electrocardiograms at rest, for long-term monitoring, during stress tests and for Holter, applying them directly to the surface of the skin. The Ag/AgCl sensor provides a stable and reliable ECG trace. The waterproof medical grade PE foam outer surface protects the skin under the electrode from air and contaminating medical fluids.

Compliant with the ANSI/AAMI EC12:2000 standard, it allows recovery of the ECG signal after defibrillator discharge. After being opened, the bag can be closed on itself in order to guarantee better conservation of the electrodes that remain inside.

The products do not contain latex.

Pre-gelled electrodes in adhesive sponge (FOAM) with Ag/AgCl sensor and liquid gel.

- Use: Adults
- Connection: clip
- Size: 36mm X 40mm (oval)
- Shape: oval

Electrodes in bag of 50 pcs.

Test result according to ANSI/AAMI EC12:2000 specification:
- Average value of the impedance measured at 10Hz for 12 pairs of electrodes: 30 Ω
- Single pair impedance: 41 Ω
- Offset Voltage: 4.3 mV
- Residual voltage after defibrillation: 12.7 mV
- Variation of polarization: 0.3 mV/s
- Impedance at 10 Hz after testing: 32 Ω
- Bias current tolerance: 16.1 mV

Maximum values according to ANSI/AAMI EC12:2000 specification:

- Average value of the impedance measured at 10Hz for 12 pairs of electrodes: 2 kΩ
- Single pair impedance: 3 kΩ
- Offset instability combined with intrinsic noise: 150 μV
- Offset Voltage: 100 mV
- Residual voltage after defibrillation: 100 mV
- Variation of polarization: 1 mV/s
- Impedance at 10 Hz after testing: 3 kΩ
- Bias current tolerance: 100 mV