Disposable electrodes for ECG, in liquid GEL with Ag/AgCl (Silver/Silver Chloride) sensor and stainless steel clip.
Disposable ECG electrodes FIAB are suitable for recording electrocardiograms at rest, for long monitoring, during stress tests and for Holter, applying them directly to the surface of the skin. The Ag/AgCl sensor provides a stable and reliable ECG tracing. The outer surface made of waterproof medical-grade PE foam material protects the skin under the electrode from air and contaminating medical fluids.
Compliant with the ANSI/AAMI EC12:2000 standard, it allows the recovery of the ECG signal after defibrillator discharge. After being opened, the bag can be closed on itself in order to ensure better preservation of the electrodes that remain inside.
The products do not contain latex.
Pre-gelled electrodes in adhesive sponge (FOAM) with Ag/AgCl sensor and liquid gel.
- Usage: Adults
- Connection: Clip
- Dimensions: 36mm X 40mm (oval)
- Shape: oval
Electrodes in a bag of 50 pcs.
Test result according to ANSI/AAMI EC12:2000 specification:
- Average impedance value measured at 10Hz for 12 pairs of electrodes: 30 Ω
- Single pair impedance: 41 Ω
- Offset Voltage: 4.3 mV
- Residual voltage after defibrillation: 12.7 mV
- Polarization variation: 0.3 mV/s
- Impedance at 10 Hz after testing: 32 Ω
- Bias current tolerance: 16.1 mV
Maximum values according to ANSI/AAMI EC12:2000 specification:
- Average impedance value measured at 10Hz for 12 pairs of electrodes: 2 kΩ
- Single pair impedance: 3 kΩ
- Offset instability combined with intrinsic noise: 150 μV
- Offset Voltage: 100 mV
- Residual voltage after defibrillation: 100 mV
- Polarization variation: 1 mV/s
- Impedance at 10 Hz after testing: 3 kΩ
- Bias current tolerance: 100 mV