VE-300 - 3-channel ECG
- Fully digital signal management
- LCD display with waveforms, configurable parameters
- High-resolution thermal printer - Built-in rechargeable Li-Ion battery for 6.5 hours of use (AC/DC powered)
- Full digital filter, avoids baseline deviation, AC, EMG interference
- Automatically adjustable base line, optimizes print position
- Selectable auto/manual function
Automatic acquisition with 12 leads, automatic ECG parameter measurement and analysis, supports an inkjet printer via USB for 12-lead printing, data storage of 120 patients and transfer via USB flashdisk, Ethernet or serial port, review and management of ECG data on Windows PC via software (optional 33336)
Internal software: GB, IT, DE, FR, ES
1 pc. Branches: standard 12 channels
Acquisition method: 12 channels simultaneously
Frequency Response: 0.05 Hz ~ 150 Hz
Sensitivity: 2.5, 5, 10, 20 (mm/mV)
Input impedance: >50 MO (10 Hz)
Filters: EMG filter 25/35/45Hz / off
Filter:
- DFT filter 0.05/0.15/0.25/ 0.32/0.5/0.67 Hz
- 150/100/75 Hz low-pass filter
- AC filter on/off
Recording Mode: Auto/Manual/Rhythm/OFF/USBPRT
Safety Standard: IEC I/CF
Power supply: AC: 100-115-220-240 V, 50/60 Hz
Built-in rechargeable lithium battery 14.8 V More than 6.5 hours, in continuous operation powered by battery
Thermal paper: 80 mm x 20 m
Print speed 5/6.25/10/12.5/25/50 mm/s (± 3%)
Dimensions: 300 x 260 x 75 mm
Weight: 2.5 kg (including battery)
Screen: LCD 192 x 64 pixels
External input/output
- Input: =100 kO; Sensitivity: 10 mm/V ± 5%
- Output: =100 kO; Sensitivity: 1 mm/V ± 5%
- Communication interface: Ethernet, RS232, USB5-lead veterinary ECG cover (Ø 4 mm, banana plug, IEC)
Veterinary clip-on electrodes (set of 5, for ECG cables Ø 3-4 mm)
1 roll of paper
Rechargeable lithium battery (2200 mAh)
Manual in: GB, IT - FR, ES, DE on request