Pre-gelled FOAM electrodes for ECG oval 36x40 mm F9089GL - Ag/AgCl - Pack of 50 electrodes
zoom_out_map
chevron_left chevron_right

Pre-gelled FOAM electrodes for ECG oval 36x40 mm F9089GL - Ag/AgCl - Pack of 50 electrodes

€4.18 Tax included

€3.43 Tax excluded

Shipping in 7/10 days.
local_shipping Shipping in 7/10 days.
Security policy
(edit with the Customer Reassurance module)
Delivery policy
(edit with the Customer Reassurance module)
Return policy
(edit with the Customer Reassurance module)
Description

Disposable ECG electrodes, in liquid GEL with Ag/AgCl (Silver/Silver Chloride) sensor and stainless steel clip.

FIAB disposable ECG electrodes are suitable for recording resting electrocardiograms, long-term monitoring, stress tests and Holter, by applying them directly to the skin surface. The Ag/AgCl sensor provides a stable and reliable ECG trace. The external surface in waterproof medical grade PE foam protects the skin under the electrode from air and contaminating medical fluids.

Compliant with ANSI/AAMI EC12:2000 standard, it allows the recovery of the ECG signal after the defibrillator discharge. The bag, after being opened, can be closed on itself in order to guarantee a better conservation of the electrodes that remain inside.

The products do not contain latex.

Pre-gelled electrodes in adhesive sponge (FOAM) with Ag/AgCl sensor and liquid gel.

- Use: Adults
- Connection: clip
- Dimensions: 36mm X 40mm (oval)
- Shape: oval

Electrodes in pack of 50 pcs.

Test results according to ANSI/AAMI EC12:2000 specification:
- Average value of impedance measured at 10Hz for 12 pairs of electrodes: 30 Ω
- Single pair impedance: 41 Ω
- Offset Voltage: 4.3mV
- Residual voltage after defibrillation: 12.7 mV
- Polarization variation: 0.3 mV/s
- Impedance at 10 Hz after testing: 32 Ω
- Bias current tolerance: 16.1 mV

Maximum values according to ANSI/AAMI EC12:2000 specification:

- Average value of impedance measured at 10Hz for 12 pairs of electrodes: 2 kΩ
- Single pair impedance: 3 kΩ
- Offset instability combined with intrinsic noise: 150 μV
- Offset Voltage: 100mV
- Residual voltage after defibrillation: 100 mV
- Polarization change: 1 mV/s
- Impedance at 10 Hz after testing: 3 kΩ
- Bias current tolerance: 100 mV

Data sheet
FIAB
ECGXF9089GL
New
8054953347141