Pre-gelled FOAM electrodes for ECG oval 36x40 mm F9089 - Ag/AgCl - Pack of 100 electrodes
zoom_out_map
chevron_left chevron_right

Pre-gelled FOAM electrodes for ECG oval 36x40 mm F9089 - Ag/AgCl - Pack of 100 electrodes

€7.49 Tax included

€6.14 Tax excluded

Shipping in 1/2 days
check Shipping in 1/2 days
Security policy
(edit with the Customer Reassurance module)
Delivery policy
(edit with the Customer Reassurance module)
Return policy
(edit with the Customer Reassurance module)
Description

Elettromedicali.it presents: Pre-gelled FOAM electrodes for oval ECG 36x40 mm F9089 - Ag/AgCl - Pack of 100 electrodes.

Discover the high quality disposable ECG electrodes from FIAB brand, available on elettromedicali.it. Designed with care and precision, these electrodes offer a reliable solution for recording the electrocardiogram.

Made with an Ag/AgCl (Silver/Silver Chloride) sensor and a stainless steel clip, FIAB disposable electrodes guarantee a stable and precise ECG tracing. Their oval shape of 36x40 mm makes them suitable for use on adults.

These electrodes are ideal for a variety of applications, from resting ECG recording to extended monitoring, from stress testing to Holter use. With their pre-gelled design and waterproof medical grade PE foam outer surface, they effectively protect the skin from airborne contaminants and medical fluids.

FIAB products comply with the ANSI/AAMI EC12:2000 standard, which allows the recovery of the ECG signal even after a defibrillator shock. Furthermore, the packaging bag, once opened, can be closed for optimal conservation of the electrodes remaining inside.

The package contains 100 electrodes, ensuring a sufficient supply for your needs.

Technical features:

- Use: Adults
- Connection: Clip
- Dimensions: 36mm X 40mm (oval)
- Shape: Oval

Test results performed according to the ANSI/AAMI EC12:2000 specification show excellent performance. The impedance measured at 10Hz for 12 pairs of electrodes is an average of 63.8 Ω, ensuring stability and accuracy in the ECG trace. In addition, the residual voltage after defibrillation is only 7.6 mV, ensuring rapid signal recovery.

The maximum values, compliant with the ANSI/AAMI EC12:2000 specification, demonstrate the high quality of these electrodes. The impedance measured at 10Hz for 12 pairs of electrodes is only 2 kΩ, while the residual voltage after defibrillation reaches only 100 mV.

Trust FIAB 's pre-gelled FOAM electrodes for oval ECG 36x40 mm F9089 - Ag/AgCl to obtain precise and reliable results in your ECG recordings. Buy now on elettromedicali.it and guarantee the best care for your patients.

CND: C020501

BD /RDM: 2251898

CLASS: |

Test results according to ANSI/AAMI EC12:2000 specification:
- Average value of impedance measured at 10Hz for 12 pairs of electrodes: 63.8 Ω
- Single pair impedance: 69 Ω
- Offset instability combined with intrinsic noise: 12.6 μV
- Offset Voltage: 0.6mV
- Residual voltage after defibrillation: 7.6 mV
- Polarization variation: 0.17 mV/s
- Impedance at 10 Hz after testing: 63.2 Ω
- Bias current tolerance: 15.2 mV

Maximum values according to ANSI/AAMI EC12:2000 specification:

- Average value of impedance measured at 10Hz for 12 pairs of electrodes: 2 kΩ
- Single pair impedance: 3 kΩ
- Offset instability combined with intrinsic noise: 150 μV
- Offset Voltage: 100mV
- Residual voltage after defibrillation: 100 mV
- Polarization change: 1 mV/s
- Impedance at 10 Hz after testing: 3 kΩ
- Bias current tolerance: 100 mV