FIRING EVO - professional electrotherapy 2 channels
Firing is a state-of-the-art portable stimulator for physiotherapy and rehabilitation.
Using the most advanced electronic technology combined with the extremely simple user interface, FIRING will satisfy the most demanding physiotherapists.
Features :D two channels completely independent: intensity and parameter controls for each channel.
20 currents/waveforms used in both rehabilitation and pain therapy.
A huge library of preset programs for most conditions.
Create and save user-defined programs and sequences.
Immediate availability of a selection of user-defined "favorite" programs.
Perineal treatments with preset programs.
Diagnostic analysis with Intensity/Time (I/T) curve.
Patient database: Stores patient data with treatment history.
Assessment of the pain level for each patient recorded at the beginning and end of treatment.
MPP system: extremely complex programs and protocols that will change frequency, pulse duration, sweep, rest/action frequency, intensity and relaxation
time. All these parameters are changed automatically during the program without operator intervention.
Possibility of connection to an ultrasound machine (optional cable required).
Large, high-quality graphic monitor for better screen visibility.
However, with all these features, FIRING is extremely easy and intuitive to use.
BENEFITS:
- improvement of the sense of balance
- learning the correct posture
- better joint stabilization through proprioceptive re-education
- better muscle coordination
Specifications:
Power supply: 115 ÷ 230 V~, 50 – 60 Hz
Power consumption: 60 VA
Weight: 3.8 Kg
Dimensions (LxWxH): 27 x 32.8 x 23 cm
Insulation class: Class I – Type BF
STIMULATION MODE:
Tens: Presents many resident programs for the most common conditions. Frequency: 2÷200Hz.
Pulse duration: 50÷600 μs. Intensity: 0÷250 mA (Ipp).
Muscle: symmetrical bidirectional impulses. It has a file with a series of programs for the treatment of different muscle conditions. Frequency: 2÷200Hz. Durability: 50÷600 μs. Intensity: 0÷250 mA (Ipp).
Bipolar interference current: bidirectional square wave current modulated in amplitude. Square wave frequency: 2000÷4000 Hz. (± 20 %). Sinusoidal modulation frequency: 5÷200 Hz. Intensity: 0÷200 mA (Ipp).
Four-pole interference: bidirectional current modulated in amplitude. Sinusoidal frequency 2000÷10,000 Hz. Sinusoidal modulation frequency 5÷200 Hz. Intensity 0÷125 mA (Ipp).
Burst: rectangular bidirectional symmetrical pulses.
Action: 0.25 s. Pause: 0.75 s. Frequency: 2÷200Hz. Pulse duration: 50÷600 μs.
Kotz: interrupted sine current.
Sine frequency: 1000÷2500 Hz. Modulation frequency: 5÷200 Hz. Intensity: 0÷250 mA (Ipp).
Trabert: rectangular, unidirectional current.
Pulse duration: 2 ms. Pause: 5 ms. Intensity: 0÷45 mA (Ip).
AMF: Rectangular bidirectional symmetrical pulses.
Increase: 2 sec. Autumn: 1 s. Rest: 2 s. Frequency: 2÷200Hz. Pulse duration: 50÷600 μs.
Faradic current: unidirectional pulses. Frequency: 1÷100 Hz. Intensity: 0÷150 mA (Ip).
Iontophoresis: interrupted unidirectional current. Frequency: 8000Hz. Intensity: 0÷50 mA (Ip).
FM: Rectangular bidirectional symmetrical pulses. Adjustable 2÷10 Hz sweep. Adjustable pulse duration. Frequency: 2÷200Hz. Pulse duration: 50÷600 μs.
Diadynamic currents: single-phase (MF), two-phase (DF), short-period (CP), long-period (LP). Intensity: 0÷50mA (ip).
Triangular/Exponential Pulses: unidirectional triangular pulses.
Duration: 1÷1000 ms. Pause: 1÷10 s. Intensity: 0÷50 mA.
Rectangular pulses: unidirectional rectangular pulses.
Duration: 1÷1000 ms. Pause: 1÷10 s. Intensity: 0÷50 mA.
Agonist/antagonist: rectangular bidirectional symmetrical pulses. Stimulation of agonist/antagonist muscles.
Agonist/antagonist with fardic current: unidirectional pulses.
Frequency: 1÷100 Hz. Intensity: 0÷150 mA (Ip).
Agonist/antagonist with Kotz current: interrupted sinusoidal current. Sine frequency: 1000÷2500 Hz. Modulation frequency: 5÷200 Hz. Intensity: 0÷250 mA (Ipp).
Diagnostics: recording of the intensity/time curve. Automatic calculation of rheobase, chronaxy, accommodation coefficient and curve drawing. Possibility to store the results and record them on databases.
Mod. High Voltage: double one-way pulse.
Polarity: positive or negative. 200 V into 5 KOhm. 160 mA into 500 Ohms.
Urology: bidirectional symmetrical pulses. Frequency: 2 ÷ 200 Hz. Pulse duration: 50÷600 μs. Maximum current: 250 mA (Ipp). Output Characteristic: Constant Voltage (CV).
Equipment:
1 pc. - FIRING EVO - electrotherapy
2 pcs. – Electrostimulation cables
4 pcs. – Electrodes 60X40 mm
2 pcs. – Electrodes 60X60 mm
2 pcs. – Electrodes 60X120 mm
4 pcs. – Bearings for EL/1
2 pcs. – Bearings for EL/2
2 pcs. – Bearings for EL/3
2 pcs. – Elastic straps 50 cm
2 pcs. – Elastic straps 100 cm
1 pc. – Black splitter cables
1 pc. – Red splitter cables
1 pc. – Black cable 2mm plug
1 pc. – Red cable 2 mm plug
1 pc. - Power cord